eIFU

MIV TMJ System

Instructions for the safe and intended use of the MIV TMJ software for CBCT-based assessment of selected temporomandibular joint parameters.

IFU rev. [XX] Document: MIVTMJ-TF-II.2 Software version: [X.X.X] Language: EN Select language version: Polish

Current regulatory status

MIV TMJ is currently undergoing preparation for the conformity assessment process under Regulation (EU) 2017/745 (MDR). The current development / research version is not intended for routine clinical use or as a basis for diagnostic or therapeutic decisions. This notice will be updated when the applicable conformity assessment process has been completed.

Controlled IFU document

Always verify that you are using the instructions corresponding to your installed software version.

01 / DOCUMENT CONTROL

Document information

This Instructions for Use document applies only to the MIV TMJ software version identified below. The user should verify the installed software version before use.

DeviceMIV TMJ System
ManufacturerMIV P.S.A.
IFU version[rev. XX]
Software version[X.X.X]
Issue date[DD.MM.YYYY]
Document IDMIVTMJ-TF-II.2
02 / DEVICE

Device description

MIV TMJ is desktop medical software designed to process CBCT imaging datasets in DICOM format and support qualified healthcare professionals in the assessment of selected anatomical and spatial parameters of the temporomandibular joints.

The system uses image-processing and AI/ML algorithms to identify predefined anatomical structures and landmarks, calculate selected parameters and present the results in a structured analysis environment.

Important

MIV TMJ is a decision-support tool. It does not provide an autonomous diagnosis and does not replace clinical examination, professional knowledge or independent clinical judgement.

03 / INTENDED PURPOSE

Intended purpose

MIV TMJ is intended to support qualified healthcare professionals in the assessment of selected parameters of the temporomandibular joints based on CBCT DICOM imaging data.

The software supports structured review of selected anatomical landmarks, reference planes, measurements and bilateral relationships of the right and left temporomandibular joints.

All automatically generated outputs must be reviewed and verified by a qualified user before being interpreted in the clinical context.

04 / USERS

Intended users

MIV TMJ is intended exclusively for qualified healthcare professionals with appropriate knowledge of craniofacial anatomy, CBCT imaging and assessment of the temporomandibular joint.

  • Dentists
  • Orthodontists
  • Prosthodontists
  • Oral and maxillofacial surgeons
  • Other appropriately trained healthcare professionals within the defined intended purpose

Professional use only

MIV TMJ is not intended for use by patients, consumers or other lay persons.

05 / PATIENT POPULATION

Patient population

MIV TMJ may be used for patients for whom a CBCT examination containing the required temporomandibular joint structures has been obtained on the basis of an independent clinical indication.

The software does not determine whether CBCT imaging should be performed and must not be used as the basis for exposing a patient to ionising radiation.

06 / CAPABILITIES

System capabilities

  • Import of CBCT DICOM imaging datasets
  • 3D visualization of selected anatomical structures
  • Automatic identification of selected anatomical landmarks
  • Segmentation of selected anatomical structures
  • Definition of selected reference planes and axes
  • Calculation of selected TMJ measurements
  • Analysis of selected radiological joint-space parameters
  • Comparison of selected right and left TMJ parameters
  • Manual review and correction of automatically generated outputs
  • Structured presentation and reporting of results
07 / INPUT DATA

CBCT input requirements

The input dataset must contain CBCT imaging data in a supported DICOM format and include the anatomical structures required for the selected analysis.

Before starting the analysis, verify:

  • correct DICOM import;
  • appropriate field of view (FOV);
  • visibility of required anatomical structures;
  • adequate image quality and spatial resolution;
  • patient positioning;
  • absence of artefacts significantly affecting interpretation;
  • correct orientation and geometry of the reconstructed volume.

WARNING – Limitations related to field of view and CBCT input quality

MIV TMJ analyses anatomical structures visible in the supplied CBCT examination. The software does not independently determine whether the CBCT examination was acquired correctly or whether the available field of view is sufficient for reliable completion of all analyses and measurements.

Before starting the analysis, the qualified user should verify image quality, field of view (FOV), patient positioning and complete visibility of the anatomical structures required for the intended analysis.

Particular attention should be paid when the anatomical structure of interest, including the temporomandibular joint, is located close to the boundary of the acquired imaging volume. Such structures may be partially truncated, incomplete or insufficiently represented for stable automated segmentation, anatomical landmark localisation or measurement geometry.

A CBCT examination may appear visually acceptable for specialist review while still containing insufficient information for reliable automated analysis.

Automatically generated segmentations, anatomical landmarks, reference planes and measurements shall be verified by the qualified user before use in clinical assessment.

If the required anatomical structures are not fully visible, or if there is any doubt regarding image quality or completeness of the input data, the automatically generated results should not be used as the sole basis for clinical decision-making.

CBCT data quality warning

If possible abnormalities in the input data are detected, patient positioning, field of view or visibility of the required anatomical structures may be insufficient for reliable analysis.

Generated results may be incomplete or inaccurate. A qualified user must verify the quality and clinical suitability of the CBCT examination before continuing.

The presence or absence of a warning does not constitute confirmation that the CBCT examination is clinically adequate or correctly acquired. Final assessment of the quality, completeness and suitability of the imaging data remains the responsibility of the qualified user.

08 / ENVIRONMENT

Hardware and use environment

MIV TMJ is intended for operation on a compatible desktop or laptop computer in a professional healthcare environment.

The computer must meet the minimum hardware and operating-system requirements specified for the installed MIV TMJ software version.

Performance

Processing time depends on the size and resolution of the CBCT dataset, the hardware configuration and the available computational resources. The user should allow the analysis to complete before reviewing the generated results.

09 / WORKFLOW

Typical analysis workflow

01

Import CBCT

Load the appropriate CBCT DICOM dataset.

02

Verify input data

Check image quality, FOV, positioning and visibility of required anatomy.

03

Automatic processing

The software processes the CBCT dataset and generates selected anatomical landmarks and structures.

04

Review AI outputs

Verify automatically generated landmarks, segmentation and anatomical references.

05

Review measurements

Assess calculated parameters and bilateral comparison of the joints.

06

Clinical interpretation

Interpret the information together with the examination, symptoms, medical history and other relevant clinical data.

10 / AI & AUTOMATION

AI-generated outputs

MIV TMJ uses automated image-analysis components, including models supporting segmentation of selected structures and identification of predefined anatomical landmarks.

The models operate as fixed inference models during normal use. Patient examinations do not automatically retrain or modify the model.

Mandatory user verification

Automatically generated landmarks, segmentations, reference planes and measurements may be incorrect. The user must review their anatomical correctness before using the generated information in further assessment.

If an automatically generated result is inconsistent with the CBCT image or cannot be reliably verified, the result must not be relied upon.

Automated analysis depends on the information actually represented in the source CBCT dataset. A structure located near the scan boundary may be only partially represented even when the examination appears visually acceptable for specialist review. This may affect segmentation, landmark localisation, reference geometry and derived measurements.

11 / MEASUREMENTS

Measurements and radiological joint space

MIV TMJ calculates selected distances, angles and spatial parameters on the basis of the CBCT dataset and verified anatomical references.

Radiological joint space

Radiological joint-space measurements describe selected distances between the bony surface of the mandibular condyle and bony structures of the articular fossa in defined directions.

Depending on the implemented analysis, these may include selected anterior, superior, posterior, medial and lateral measurements.

Interpretation limitation

Radiological joint space is an imaging measurement of bony spatial relationships. It does not directly visualise the articular disc or other soft tissues.

Its interpretation may also be affected by the mandibular position at the time of image acquisition.

12 / SAFETY

Warnings and limitations

  • Do not use an automatically generated result without verifying it against the source CBCT data.
  • Do not use MIV TMJ as the sole basis for diagnosis or therapeutic decisions.
  • Do not assume that bilateral symmetry indicates normal anatomy or normal function.
  • A static CBCT examination does not by itself provide a complete assessment of TMJ function.
  • CBCT primarily provides information about mineralised structures. Soft-tissue assessment may require other diagnostic methods.
  • Image artefacts, incomplete FOV, incorrect patient positioning, unusual anatomy or structures located near the boundary of the scan may reduce the reliability of automated outputs.
  • The software does not independently confirm that the examination was correctly acquired or that the FOV is sufficient for every intended analysis.
  • A CBCT examination may appear visually acceptable for specialist review while still containing insufficient information for reliable automated analysis.
  • The absence of an input-quality warning does not confirm that the examination is clinically adequate or complete.
  • Zooming or enlarging an image does not increase the underlying information or spatial resolution of the original CBCT acquisition.
  • Results must always be interpreted together with relevant clinical information.
13 / SMALL FOV

Examinations with a limited field of view

A limited FOV does not automatically make a CBCT examination unsuitable for assessment of the TMJ.

If the required TMJ structures are fully visible and sufficiently represented, the examination may allow assessment of their visible bony anatomy.

Scan-boundary limitation

Particular attention is required when the temporomandibular joint or another structure required for the analysis is located at or close to the boundary of the acquired volume. The structure may be partially truncated, incomplete or insufficiently represented for stable automated segmentation, landmark localisation or measurement geometry.

The examination may still be visually interpretable by a specialist, but this does not guarantee that it contains sufficient data for reliable automated analysis. The qualified user must verify completeness of the required anatomy before relying on automatically generated results.

Topographical limitation

If the field of view does not allow the user to verify the relationship of the mandible to the maxilla or the patient's occlusal position during acquisition, anatomical structures may still be assessed, but interpretation of their topographical relationships must be limited accordingly.

14 / DATA SECURITY

Patient data and cybersecurity

Users must process patient data in accordance with applicable data-protection requirements and the policies of their healthcare organisation.

  • Use individual user credentials.
  • Do not share passwords.
  • Use supported versions of the software.
  • Install security and software updates provided by MIV P.S.A.
  • Protect the workstation against unauthorised physical and remote access.
  • Do not modify application files, AI models or software dependencies.
15 / SOFTWARE UPDATES

Updates and version control

MIV TMJ software versions are released through a controlled software release process.

Users should only use versions supported by MIV P.S.A. and should ensure that the IFU corresponds to the installed software version.

AI model versions

Changes to an AI model, its preprocessing, postprocessing or execution environment are subject to controlled verification and validation before release.

16 / TROUBLESHOOTING

Troubleshooting

SituationRecommended action
DICOM dataset cannot be importedVerify file integrity, DICOM compatibility and completeness of the imaging dataset.
CBCT quality warning displayedReview positioning, FOV, image quality, artefacts and visibility of the required structures. Cancel the analysis if the data are not clinically suitable.
TMJ or required anatomy is close to the scan boundaryVerify that the structure is fully represented in the acquired volume. If it is partially truncated, incomplete or insufficiently visible, do not rely on the automated analysis as the sole basis for clinical decision-making.
No warning is displayed, but image quality or FOV is questionableDo not interpret the absence of a warning as confirmation of scan adequacy. Independently verify image quality, FOV and visibility of the required anatomy.
Automatic landmark is incorrectly positionedVerify the landmark against the CBCT image and correct it manually if the function is available. Do not accept an incorrect result.
Segmentation appears anatomically incorrectDo not rely on measurements derived from the incorrect segmentation. Review the source image and repeat or discontinue the analysis.
Processing does not completeDo not interpret incomplete outputs. Restart the workflow according to the application instructions or contact support.
Unexpected or implausible measurementVerify the source image, landmark positions, segmentation and reference planes before interpreting the measurement.
17 / MANUFACTURER

Manufacturer and support

MIV P.S.A.
Ignacego Mościckiego 1
24-110 Puławy
Poland

E-mail: info@mivtmj.com

Website: mivtmj.com
Current online IFU: mivtmj.com/ifu/
User manual: mivtmj.com/manual/

A paper or electronic copy of the current Instructions for Use may be requested from MIV P.S.A. using the contact details above.

MIV P.S.A.

Ignacego Mościckiego 1, 24-110 Puławy, Poland

info@mivtmj.com